Failure Description | Possible Cause | Troubleshooting |
---|---|---|
External module cannot be loaded | Defective parameter module | Replace the malfunctioning parameter module with a known good module. If the patient monitor identifies the replacement module and can start measurement, it indicates that the original module is faulty. |
Power supply to the internal module rack is defective | The internal module rack backpane or main control board may fail, and 12 V cannot be output. | |
Unable to recognize modules | Cable defective from the main control board to the internal module rack backpane |
|
Version inconsistency between the module communication board and the internal module rack backpane | Upgrade the module or internal module rack backplane software. | |
The main board is faulty. | Replace the main board. | |
Source: Page 5-6 Mindray ePM Service Manual |