Failure Description | Possible Cause | Troubleshooting |
---|---|---|
Alarm LED off or cannot be turned off while the audible alarm is provided | Cable defective or improperly connected |
|
Alarm lamp board failure | Replace the alarm lamp board. | |
The main board is faulty. | Replace the main board. | |
No audible alarm sounds emitted when alarm lamp is activated | Audible alarm disabled | Check whether the alarm sound is muted; select Main Menu → System >> → Maintenance >> → enter the user maintenance password → Alarm >>, and adjust the Minimum Alarm Volume to an appropriate value in the pop-up menu. In the Others window of the Alarm Setup menu, set Alm Volume to appropriate setting. |
Speaker failure | Replace the speaker. | |
Cable defective or improperly connected | Verify the cable connecting the speaker to the main control board is properly connected. | |
The main board is faulty. | Replace the main board. | |
Source: Figure 5.5.3 Page 5-7 Mindray ePM Service Manual |