Failure Description | Possible Cause | Troubleshooting |
---|---|---|
ECG/Resp/NIBP/Temp/IBP failure | Incorrect software version | Check the MPM and system software versions and update the software. |
Parameter circuit is damaged | Replace the main board. | |
Cable defective or improperly connected |
| |
Accessories may be damaged | Replace the accessories. | |
Check whether the parameter configuration is correct | According to the user manual, check whether the corresponding parameter configuration is enabled | |
SPO2 module (Masimo/Nellcor) failure | Accessories may be damaged | Replace the accessories. |
SPO2 module failure | Replace the module. | |
CO2 (built-in) module failure | Internal hose connection failure |
|
System software configuration error | Check if the modules in the system software are configured correctly. | |
Module damaged | Replace the module. | |
Cable defective or improperly connected |
| |
The main board is faulty. | Replace the main board. | |
C.O. (built-in) module failure | System software configuration error | Check if the modules in the system software are configured correctly. |
Module damaged | Replace the module. | |
Line connection failure |
| |
The main board is faulty. | Replace the main board. | |
CO2/AG/IBP/C.O. single- or dual-slot modules failure | Module damaged | Replace the module. |
Internal module rack backpane defective | Replace the backboard. | |
Incorrect firmware version of the integral module rack | Check the firmware version and update. | |
Cable defective or improperly connected |
| |
Source: 5.5.8 Page 5-9, 5-10 Mindray ePM Service Manual |