The ePM 10M/12M/10A/12MA uses:
Masimo LNCS DCI Adult (RD SET)
Part Number: 040-003376-00
Alternate Part Number: 92-4050
SpO2 Specifications for ePM Montiors | ||
---|---|---|
Alarm Limit | Range (%) | Step (%) |
SpO2 High | (low limit + 2) to 100 | 1 |
SpO2 Low | Masimo: (Desat+1) to (high limit - 2) Nellcor: (Desat+1) or 20 (whichever is greater) to (high limit - 2) | |
SpO2 Desat Low | 0 to (SpO2 low limit - 1) | |
Source: A-10 Mindray ePM Operators Manual |
Masimo SpO2 Module | |
---|---|
Standards | |
Measurement Range | meets the requirements of ISO 80601-2-61: 2011 |
Resolution | 1 to 100% |
Response Time | 1% |
Accuracy1 | 70 to 100%: ±2% (measured without motion in adult/pediatric mode) |
Refresh Rate | ≤ 1 s |
SpO2 averaging time | 2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s |
Low perfusion conditions | Pulse amplitude: >0.02% Light penetration: >5% |
Low perfusion SpO2 accuracy2 | ±2% |
PI measurement range | 0.02 to 20% |
1 The Masimo pulse oximeter with sensors have been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. One percent was added to the accuracies for neonatal sensors to account for accuracy variation due to properties of fetal hemoglobin. 2 The Masimo pulse oximeter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. | |
Source: A-10 Mindray ePM Operators Manual |