To ensure the correct functioning of the ENBIO S sterilizer, it is recommended to perform periodic service inspections and replace parts subject to wearing down in accordance with the following schedule and periodic inspection of individual sterilizer elements in accordance with the following guidelines.
Name |
Replacement frequency |
---|---|
HEPA bacteriological filter |
Every 1000 cycles or every 12 months |
Connection/water removal tube |
If damage is observed or once a year |
Plugs for water/condensate containers |
If damage is observed |
Element subject to inspection |
Inspection frequency |
Performed By |
---|---|---|
Front seal |
weekly or in the event of incorrect functioning |
User |
Bacteriological filter |
every week |
User |
Connection / water removal tube |
weekly or in the event of incorrect operation |
User |
Container plugs |
weekly |
User |
What are common replacement items needed during periodic inspections?
Name |
Part number |
---|---|
Front seal |
ST1-UL1 |
Bacteriological filter |
DZ0035 |
Connection/water supply tube |
ST1-HW1 |
Connection/water supply tube |
ST1-HW2 |
Rubber plug for the water container |
ST1-KS1 |
Rubber plug for the condensate container |
ST1-KS2 |