DOOR OPEN message displayed on screen
Source | Cause | Solution | P/N |
|
---|---|---|---|---|
Door reed switch magnetic actuator | Weak magnet, loss of magnetic properties | Replace front door | 1-8-14672F2 | |
Reed switch | Faulty reed switch, or loose connector | Replace front door | 1-8-14672F2 | |
Main control board | Lack of signal | Replace the main control board. To obtain the correct version of the board, please contact service and provide: photo of the board, device S/N and process count (if you cannot read this on the device, get the data from the last log in Enbio Data Viewer) | 1-8-25103E2 |