All leaks should be corrected before proceeding with troubleshooting
Cycle Fault #1 The cassette temperature failed to reach 95°C within a time-out period.
An extremely large steam leak (Statim 5000). Replace cassette seal, lid or tray. End user repair
An extremely large load (Statim 5000). End user repair
If the boiler does not heat up, check the thermal fuse, using an ohmmeter to check for continuity between J1-3 & lower terminal on the boiler. If you read less than 1Ω thermal fuse is good. If the thermal fuse checks good, proceed to step 3; if bad, replace and check the water pump using the Pump Tester Bottle. Field Service Tech. repair
Check resistance reading between J1-3 & J1-4, should read approx. 11Ω. Field Service Tech. repair
Check for line voltage at terminals J1-3 & J1-4 during warm-up. If the reading is good, proceed to step 4, if bad, PCB (Printed Circuit Board) is defective. Shop repair
Check for line voltage at boiler terminals. Field Service Tech. repair