The cassette temperature has dropped below a set point. If the cassette can be removed normally after venting:
Check for visible steam leaks from the cassette; if the cassette is leaking, repair it as needed. Replace the cassette seal, lid, or tray. End user repair
Check the solenoid and make sure the plunger is not sticking. Field Service Tech. repair
Verify that the check valve and pressure relief valves are not leaking. Field Service Tech. repair
If the cassette is hard to remove after venting: (Statim 2000 only)
Check for kinked or pinched exhaust tubing. End user repair
Check for a clogged venturi in the left rear of the cassette tray. Clean as needed. End user repair
The solenoid valve is failing to open. Disassemble and check that plunger slides smoothly into the plunger tube. Field Service Tech repair