During conditioning, the cassette temperature has failed to drop to 115°C during the Unwrapped or Wrapped Cycle, or the temperature has failed to drop to 110°C during the Rubber and Plastics Cycle in the purge conditioning stage.
Check for kinked or pinched exhaust tubing. End user repair
Check for a clogged venturi in the left rear of the cassette tray. Clean as needed. End user repair
Listen for normal clicking sounds when the cassette is inserted and removed. If there is no clicking sound, the solenoid valve fails to open. Disassemble and check that plunger slides smoothly into the plunger tube. Field Service Tech repair
If the Statim exhaust tubing is connected to a VistaCool or StatFlow system, proceed as follows.
Ensure the cold water supply to the water tank is turned ON. End user repair
Ensure the water tank is completely filled. Field Service Tech repair
Check the brass water valve on top of the water tank is operating properly. Insert a small screwdriver in the view port hole on the brass stem of the thermal regulator on top of the tank and push up on the water piston. The piston operates when water can be seen leaving the ½” overflow fitting on the tank and running to the drain adapter. Field Service Tech repair
Check the in-line thermal sensor on the manifold drain line is open. Field Service Tech repair