The sterilization phase failed to start within 3 minutes of the cassette reaching the sterilization temperature. Note: the unit has to fail on three consecutive cycles for CF26 to display. “Cycle Interrupted” displayed for the first two occurrences.
Check for visible steam leaks from the cassette. If the cassette leaks, repair it as needed, and replace the cassette seal, lid, or tray. End user repair
Check the solenoid for debris and make sure the plunger is not sticking. Field Service Tech. repair
Calibrate Validation thermocouple. Field Service Tech. repair