Communication error, the LCD controller PCB failed to receive messages from the Rev. 7 PCB.
Check that 5V is available on Rev. 7 PCB. Greenlight is lit on the front of the PCB. Field Service Tech. repair
Check all fuses on Rev. 7 PCB. Field Service Tech. repair
Check all cables for a bad connection. Field Service Tech. repair