Check air filters; if dirty, replace them. End user repair
Ensure that the cassette is clean. End user repair
The drain tube must run directly to the condenser bottle with no dips, loops, or kinks. End user repair
Do not stack wraps. End user repair
Invert mesh rack to provide air space below wraps. End user repair
Set the level in the user menu so the bubble is green. End user repair
Check for airflow through the unit. While the Statim is running in the drying cycle, remove exhaust tubing from the top of the waste bottle (be careful tubing may be hot). Place tubing into a cup of water, vigorous bubbles should appear in the cup water. If air bubbles do not appear, check airflow from the compressor to the waste bottle. End-user or Field Service Tech. repair