This procedure applies to the mineral-free water reservoir (left) and the waste-water reservoir (on the right).
Note: Improper Water level icon appears when the water reservoir needs
to be filled. (When there is a need to drain the wake reservoir – the message
will be displayed) .
The drain valves are located on the front right side of the autoclave after the
the door is opened.
To drain the reservoir:
See item (5) with the plastic hose (6) attached to it (supplied with the
autoclave).Put the other end of the hose into a drain bucket.
Insert part (5) into valve (3) and press it until you hear a click. The drain valve opens immediately.
When the water reservoir is empty, press part (4). Item (5) will pop out approx. 3mm, and the drain valve will be closed. Remove item (6) with the plastic tube.