Before proceeding, make sure that the electric cord is disconnected and that there is no pressure in the chamber.
Use scissors to open the filter bag and not sharp blades or pointed instruments.
Check that the new filter has not exceeded the maximum shelf life.
Carefully un-pack the new filter and examine it for any signs of damage.
Remove any protective packaging before inserting the filter into place.
Carefully Insert the new filter into the housing. Do not force.
The AIR filter is located mounted under the cover (see below)
Unscrew the filter and Pull the old filter out (see below).
Disconnect the pipe.
Assemble the new filter, and ensure the new filter is all the way in and seated properly.
Ensure the new filter is all the way in and seated properly.
Perform the testing process procedure.
Note: Make sure that the arrow on the filter body points inwards toward the
chamber. Make sure that you don’t bend the filter pipe when reattaching it.
Note: Replace the air filter every 6 months or after 1000 cycles (whichever is
the shorter period).